Programmer’s Guide System overview 7
Data transfer and storage within the Acuity System
network
The Acuity System stores patient data for variable amounts of time, depending upon
Acuity System license configuration.
Data export from patient monitors to the Acuity System
Welch Allyn portable patient monitors store vital-sign data and send it to the Acuity
System in the following ways:
Table 1. Patient monitor data storage and export to the Acuity System
Patient Monitor Means of data export to
Acuity System
Mode of data
export to Acuity
System during
normal conditions
Patient monitor
data storage
during temporary
disconnection or
dropout
Data export to
Acuity System
after temporary
disconnection or
dropout
Hardwired monitors:
Propaq
®
Encore,
Propaq CS
Unshielded twisted pair
(UTP)
Stream Data stored up to
monitor storage
capacity
Trends data only,
sent in batch mode
Hardwired Welch
Allyn 1500 Patient
Monitor
Ethernet Stream Data stored up to 24
hours
Trends data only,
sent in batch mode
Wireless monitors:
Propaq CS, Propaq LT
and Micropaq
®
802.11 wireless LAN Stream Propaqs: Data stored
up to monitor
storage capacity
Micropaq: No
storage
Propaqs: Trends data
only, sent in batch
mode
Micropaq: data lost