Directions for Use Chapter 3 Calibrating the Spirometer 27
If desired, you can press Stop any time. Softkeys will change, as described in Step 11
on page 28.
Otherwise, when no air has moved for three seconds, the following happens:
For verifications or single-flow calibrations
The results display.
For multiple-flow calibrations
Another simulated exhalation screen appears. Repeat from Step 8 twice more.
The results display.
10. Review your results.
Check the error percentages for the expired and inspired volumes. Both/all must be
less than ±3% for your calibration to be acceptable. For single-flow calibrations, the
measured and adjusted curves should match. See the following examples.
Figure 14. Single-Flow, Poor Results Figure 15. Single-Flow, Good Results
Figure 16. Multiple-Flow, Poor Results Figure 17. Multiple-Flow, Good Results
Calibration Results
9:17AM Oct 16 08
Syringe Volume: 2.998 L
Expired Volume: 2.834 L (-5.5%)
Inspired Volume: -2.906 L (-3.1 %)
Temperature: 74.0 F
Humidity: 57.0 %
Pressure: 739.4 mmHg
I-Gain: 1.19
E-Gain: 1.09
Legend
Measured Adjusted
AcceptRetry Exit
Vol (L)
Time (s)
Errors > ± 3%. Curves do not match. Calibration is
not acceptable.
Calibration Results
9:17AM Oct 16 08
Syringe Volume: 2.998 L
Expired Volume: 3.010 L (0.4%)
Inspired Volume: -2.936 L (-2.1%)
Temperature: 74.0 F
Humidity: 57.0 %
Pressure: 739.4 mmHg
I-Gain: 1.17
E-Gain: 0.99
Legend
Measured Adjusted
AcceptRetry Exit
Vol (L)
Time (s)
Both errors < ± 3%. Curves match. Calibration is
acceptable.
Calibration Failed
9:17AM Oct 16 08
Syringe Volume: 2.998 L
Expired Volume
0.5 L/s: 2.83 (-5.8%)
1.0 L/s: 2.98 (-0.7%)
3.0 L/s: 2.81 (-6.2%)
Temperature: 74.0 F
Humidity: 57.0 %
Pressure: 739.4 mmHg
I-Gain: 1.22
E-Gain: 1.06
Legend
Measured Adjusted
Retry Exit
Vol (L)
Time (s)
Errors > ± 3%. Calibration is not acceptable.
Calibration Results
9:17AM Oct 16 08
Syringe Volume: 2.998 L
Expired Volume
0.5 L/s: 3.05 (1.6%)
1.0 L/s: 2.98 (-0.6%)
3.0 L/s: 3.05 (1.6%)
Temperature: 74.0 F
Humidity: 57.0 %
Pressure: 739.4 mmHg
I-Gain: 1.19
E-Gain: 1.02
Legend
Measured Adjusted
Retry Exit
Vol (L)
Time (s)
Accept
All errors < ± 3%. Calibration is acceptable.