Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, andAccessoriesCleaning, Disinfection, and Sterilization InstructionsMat’l 103251 Ver. B(Thi
8 ENGLISH Cleaning, Disinfection, and Sterilization InstructionsPower SupplyThe following cleaning instructions apply to the Power Cord ONLY:1. Discon
Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 9Rechargeable HandleCleaning1. Using CaviWipes, wipe the surface to
10 ENGLISH Cleaning, Disinfection, and Sterilization Instructions
ii Cleaning, Disinfection, and Sterilization Instructions
Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1ENGLISHIntended UseThe Welch Allyn Rigid Reusable & Single Use
2 ENGLISH Cleaning, Disinfection, and Sterilization InstructionsRigid Sigmoidoscopes, Anoscopes, and Light Carrier*Suction Tubes not illustratedCleani
Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 34. Allow the item to dry for 24 hours or until no moisture remains
4 ENGLISH Cleaning, Disinfection, and Sterilization Instructions4. After performing a reprocessing procedure, wipe the lamp with a soft, dry, lint fre
Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 5Insufflation SystemCaution Welch Allyn recommends the use of the S
6 ENGLISH Cleaning, Disinfection, and Sterilization InstructionsCleaning1. Prepare an enzymatic detergent formulated for endoscopic instruments accord
Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 7Sterilization1. Follow steps 1 through 10 of the Cleaning instruct