Technical Manual
5-4 Welch-Allyn, Inc.
70898-0000A
Troubleshooting Table (continued)
SYMPTOM POSSIBLE CAUSE PROCEDURE
Display problems
(continued)
Cracked LCD Inspect LCD for hairline cracks, especially
in corners.
Microprocessor failure Check for improper soldering of pins,
crystal operation on O-scope, proper reset.
No beeper sound Defective horn Replace horn.
Broken connection Check continuity from U1-49 to horn pin 2
and from ground to horn pin 1.
Defective U1 Check for signal with O-scope at U1-49.
Replace microcontroller U1.
No Timer function Defective Timer button
S1
Check for low level (gnd.) signal at U1-26
when timer button S1 is pressed. Replace
button if signal is high.
Broken trace Check for low level (gnd.) signal at U1-26
when timer button S1 is pressed. Replace
Button if signal is high.
No Backlight Defective backlight Check LED D3 and resistor R23.
No /BLGTHCTL signal Check that signal at U1-42 goes low when
unit is in on.
Recalled temperature is
not correct.
Unit switched to Monitor
mode
If unit is in Monitor mode (whether by the
user switch or automatically), the stored
temperature is the last one seen by the
instrument. This is usually lower than the
patient temperature since the probe drops
in temperature after removal from the
patient.
Probe: Wrong type
displayed
Missing Vcc power to
R10 and/or R11
With no probe installed, check that probe
connector pins J1-B and E are both pulled
high when any function is active (recall or
timer).