Troubleshooting
SureTemp
®
Model 678/SureTemp
®
Model 679 5-5
Troubleshooting Table (continued)
SYMPTOM POSSIBLE CAUSE PROCEDURE
Probe: Wrong type
displayed (continued)
Incorrect wiring of
probe
Oral probes should have a short between
pins B, E and F (refer to instrument PCB
designators for probe pin definition).
Rectal probes should have a short between
pins E and F but open between pins B and
E or F.
Replace with new probe.
Normal/Monitor Mode
switching problems
Ambient above 33.9
°
C
(93.0
°
F)
Causes auto switch to Monitor mode.
Switched to Monitor
mode before probe in
mouth
If 60 seconds pass after ready in Normal
mode, unit switches to Monitor mode.
If probe is still cooling from a previous
temperature and used immediately, it might
sense ambient to be above 33.9
°
C
(93.0
°
F).
Defective Mode button Check mode button for proper function.
Check U1-25 for low level signal when
button is pressed.
Deffective Probe Replace probe.
Cannot enter Biotech
Mode
Mode button not
pressed, and or
Instrument not in wall
holder (678 only)
Mode button must be pressed while
instrument is in the wall holder.
Probe not connected
and or probe shaft not
inserted and removed
from probe well
Must connect probe to instrument and
remove the probe shaft from the probe well
while the instrument is in wall holder and
mode button is pressed.
Failed component,
broken trace
Check proper Mode button, Probe switch
and security switch operation.