Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details Host
controller
action
F),
respectively.
If conditions
are valid and
the problem
persists,
replace the
probe. If
problem
persists,
replace
module.
8 3
Internal calibration
resistor (RCAL) on the
board is damaged or
contaminated (pulse too
long).
Module/
Equipment alert
47 E4.1 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 4
Internal calibration
resistor (RCAL) on the
board is damaged or
contaminated (pulse too
short).
Module/
Equipment alert
48 E4.1 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 5 Internal circuit
validation resistor (PTB)
on the board is damaged
(value is over).
Module/
Equipment alert
49 E4.0 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 6 Internal circuit
validation resistor (PTB)
on the board is damaged
(value is under).
Module/
Equipment alert
50 E4.0 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 7 A/D measurement timed
out.
Module/
Equipment alert
51 E0.2 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 8
The probe was not
characterized/calibrated.
Probe/ Equipment
alert
52 E0.9 Probe
malfunction;
replace probe.
Indicate
probe
error