Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details Host
controller
action
problem
persists,
replace the
module.
8 25 Heater indicates off
when turned on.
Module/
Equipment alert
81
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Indicate
module
error
8 26 HTR_Q on and HTRC is
off, but still has voltage.
Module/
Equipment alert
82
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 27 HTR_Q is tri-stated with
HTRC enabled and has
heater power.
Module/
Equipment alert
83
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 28 Turned Q&C on and
heater voltage not high
enough.
Module/
Equipment alert
84
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Indicate
module
error
8 29 Heater hardware failsafe
should have turned off
but did not.
Module/
Equipment alert
85 E5.2
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 30 Probe over 112° F /
43.3° C.
Module/
Equipment alert
21
E0.4 Probe tip over
temperature.
Heater error or
probe
construction
problem.
Assure that
module is
within
ambient temp
limits. If the
problem
Power
down
Komentarze do niniejszej Instrukcji