Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details Host
controller
action
persists,
replace the
probe.
8 31
Excessive heater energy.
Probe/ Equipment
alert
22
E0.1
Probe warmer
is not
working. If the
problem
persists,
replace the
probe.
Power
down
8 32
Host controller interface
error.
Module/ Host
controller alert
Internal error Internal error.
Try another
predict cycle.
If the problem
persists,
replace the
module.
Clear
8 33
Over ambient
temperature.
Ambient/
Equipment alert
24 Device
environment is
too warm.
Move module
to within
ambient
temperature
range.
Clear
8 34 Under ambient
temperature.
Ambient/
Equipment alert
25 Device
environment is
too cold.
Move module
to within
ambient
temperature
range.
Clear
8 35 Invalid SureTemp Plus
algorithm.
Module/ Host
controller alert
Internal Error Internal error.
Try another
predict cycle.
If the problem
persists,
replace the
module.
Indicate
module
error
8 36 Over maximum battery
volts.
Power/ Host
controller alert
59 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
Komentarze do niniejszej Instrukcji