Welch-allyn SureTemp PlusModule - User Manual Instrukcja Użytkownika Strona 36

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 85
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 35
Packet Structure SureTemp Plus Module OEM Implementation
36 Confidential Welch Allyn OEM Technologies
TEMP (19 data bytes)
Type Description Values/Units
float32 “Final/Predicted” temperature in °Kelvin
float32 Current Probe Temperature in °Kelvin
uint32 Sample Counter - 173 mS/sample
int8 Method
0: Normal
1: Monitor
2: Clinical
int8 Algorithm
0: Oral
1: Rectal
2: Pediatric Axillary
3: Cal Key
4: Adult Axillary
int8 SureTemp state
0: Idle
1: Not Ready
2: Ready
3: Predict
4: Predict display
5: Predict position
6: Final
7: Final position
8: Final over temp
9: Final over temp position
10: Final under temp
11: Final under temp position
12: No predict (monitor mode)
13: End normal status
20: Error status
int16 Error code
0: No error
See Error Codes (page 65)
boolean active Probe out and Active
int8 Physical probe type connected
0: Oral/Axillary
1: Cal Key
2: Rectal
3: No Probe
Przeglądanie stron 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 84 85

Komentarze do niniejszej Instrukcji

Brak uwag