Welch-allyn SureTemp PlusModule - User Manual Instrukcja Użytkownika Strona 79

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 85
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 78
SureTemp Plus Module OEM Implementation Developer’s Kit
Welch Allyn OEM Technologies Confidential 79
Normal Mode
1. The program defaults to Normal (predict) mode upon initial operation; however, the mode
settings are remembered. With the probe in the well, select the Mode item from the menu
bar. Verify that none of the three mode choices is selected from the drop-down menu list.
The unit is now ready to display temperatures in Normal mode when the probe is
removed from the probe well.
2. The thermometer remains in Normal mode until another mode is selected (checked).
Monitor Mode
1. With the probe removed from the well, use the mouse to click on the Mode button on the
Handheld.
2. The thermometer remains in Monitor mode until the Monitor menu item is disabled
(unchecked).
note Use only Welch Allyn probe covers. The use of other manufacturer’s probe covers or no
probe cover may produce temperature measurement errors and/or inaccuracy.
note When using the Monitor mode, allow the temperature readout to stabilize for 3 minutes
for oral and rectal temperatures and 5 minutes for axillary temperatures.
Przeglądanie stron 78
1 2 ... 74 75 76 77 78 79 80 81 82 83 84 85

Komentarze do niniejszej Instrukcji

Brak uwag