SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 45
(3101) Send Digital Input Status
Causes the module to send the status of all the device’s digital inputs to the host. This is a
short representation of all of the switches and buttons on the device. 1 is pressed or closed,
and 0 is released, or open.
Command Code (hex)
3101
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
2 uint16 Each bit represents a different button or switch on
the device.
15: Reserved
14: Reserved
13: Reserved
12: Reserved
11: Reserved
10: Probe switch
09: Probe attached
08: Probe well inserted
07: Spare
06: Spare
05: Spare
04: Spare
03: Spare
02: Spare
01: Spare
00: Spare